From Pshtaz Teb Zaman diagnostic.\
Package size: 96 tests.
No. of test:96 tests.
Linearity:
|
|
|
Recovery
(%) | ||||
|
No. |
CEA
(ng/ml) undiluted undiluted specimen |
1:2 |
1:4 |
1:8 |
1:16 |
1:32 |
|
1 |
84 |
104 |
98 |
89 |
106 |
93 |
|
2 |
57 |
99 |
100 |
92 |
95 |
90 |
|
3 |
38 |
102 |
95 |
98 |
110 |
108 |
Precision:
Intra-assay:
|
No |
No. of
tests performed |
Mean
(ng/ml) |
SD
(ng/ml) |
CV % |
|
1 |
24 |
4.5 |
0.15 |
3.3 |
|
2 |
24 |
23.2 |
0.96 |
4.1 |
|
3 |
24 |
67.3 |
2.50 |
3.7 |
Inter-assay:
|
No |
No. of
tests performed |
Mean
(ng/ml) |
SD
(ng/ml) |
CV % |
|
1 |
10 |
5.5 |
0.28 |
5.1 |
|
2 |
10 |
22.2 |
1.19 |
5.3 |
|
3 |
10 |
74.0 |
5.10 |
6.9 |
From J.Mitra &Co. Put. Ltd.
Principle: Based on competitive immunoassay
quantitative detection of Vitamin B12 (cyanocobalamin) in human serum.
Excellent sensitivity & specificity.
Linearity Range: 50- 2000 pg/ml.
Analytical Sensitivity: 50 pg/ml.
Breakaway microwell strips.
Color coded reagents to monitor procedural steps.
Assay procedure: 125 mins.
Prompt service back-up.
Longer shelf life:18 Months at 2-8°C.
Convenient pack size: 96 tests.
From J.Mitra &Co. Put. Ltd.
Principle: Based on competitive immunoassay
quantitative detection of Total Thyroxine (T4) in human Serum / Plasma.
Linearity Range: 0.4 µg/dl to 25 µg/dl.
Analytical Sensitivity 0.4 µg/dl.
Breakaway microwell strips.
Color coded reagents to monitor procedural steps.
Assay procedure: 75 mins.
Prompt service back-up.
Longer shelf life: 18 Months at 2-8°C.
Convenient pack size:24 tests and 96 tests.
From J.Mitra &Co. Put. Ltd.
Principle: Based on competitive immunoassay.
Quantitative detection of Total Triidothyronin (T3) in Human Serum / Plasma.
Linearity Range: 0.1 ng/ml to 7.5 ng/ml.
Analytical Sensitivity 0.1 ng/ ml.
Breakaway microwell strips.
Color coded reagents to monitor procedural steps.
Assay procedure: 75 mins.
Prompt service back-up.
Longer shelf life: 18 Months at 2-8°C.
Convenient pack size: 96 tests.
From J.Mitra &Co. Put. Ltd.
Detects all the 11 subtype of HBsAg.
Antigen Sensitivity- 0.5 ng/ml.
No sample preparation required.
See through Device for easy result interpretation.
Shelf life-24 months at 2-30°C.
100% Sensitivity & 98.0% Specificity by WHO Evaluation.
20 Test Pack.
From J.Mitra &Co. Put. Ltd.
Use of Envelope Antigens gp41,C terminus of gp120 for HIV-1& gp36 for HIV-2.
Detection of Group O & subtype C for HIV-1 prevalent in India.
Color coded reagents to monitor procedural steps.
Assay procedure: 90mins.
Longer shelf life: 15 months at 2-8°C.
Sensitivity & Specificity of 100% by WHO.
From J.Mitra &Co. Put. Ltd.
Detects all the known 11-subtypes of HBV.
Antigen Sensitivity-0.1ng/ml.
Two step Immunoassay.
Short protocol: Based on single washing step.
Assay Procedure: 90 mins.
Shelf life-15 months at 2-8°C.
Convenient pack size :96 tests.
Break-away wells for minimal wastage.
Easy interpretation of result.
Sensitivity & Specificity 100% by WHO.
From J.Mitra &Co. Put. Ltd.
Results within 3 minutes.
100% Sensitivity & 98.9% Specificity as per WHO evaluation.
50 Test Pack.
From J.Mitra &Co. Put. Ltd.
Rapid ,visual 4th generation test.
Unique Combination of HCV Antigens Core, NS3, NS4 & NS5.
Based on Flow Through Technology.
Use of unique combination of Core, NS3, NS4, NS5 antigen.
Detects all the subtypes of HCV.
Result with in 3 min.
Shelf life : 15 months at 2-8°C.
Sensitivity- 100 %.
Pecificity- 99.8 %.
From J.Mitra &Co. Put. Ltd.
3rd Generation Assay.
Unique combination of HCV antigens for E1, E2, Core, NS3, NS4 & NS5.
Break-away microwell strips.
Based on Indirect Elisa principle.
Colour coded reagents to monitor procedural steps.
Short assay procedure: 90mins.
Longer shelf life: 15 months at 2-8°C.
Sensitivity 100% & Specificity 97.4% by WHO.
From labcare diagnostic.
Endpoint.
Package size: 50x1.
No. of tests: 50.
Linearity: 180 mEq/l.
From labcare diagnostic.
Fixed time.
Package size: 1x50 tests.
No. of tests: 50 tests.
Linearity: THE Linearity limit and the measurement range depends on the sample volume, although the sensitivity of the test will proportionally decrease.
Sensitivity: 3.34 Ma.. IU /MlL.
Precision:
|
|
Intra-assay
(n=10) | ||
|
Mean
(mg/dl) |
135 |
236 |
372 |
|
SD |
3.4 |
16.2 |
5.9 |
|
CV |
2.5 |
2.3 |
1.6 |
|
|
Inter-assay
(n=20) | ||
|
Mean
(mg/dl) |
135 |
236 |
372 |
|
SD |
7.9 |
13.2 |
17.8 |
|
CV |
5.9 |
5.5 |
4.8 |
From labcare diagnostic.
Endpoint.
Package size: 1x15 tests.
No. of tests: 15 tests.
Linearity: 7.0 mEq/l.
From labcare diagnostic.
Fixed time.
Package size:1x50 tests.
No of tests: 50 tests.
Linearity: The linearity limit depends on the sample reagent ratio, as well as the analyzer used.
Sensitivity: Δ 3.8 mA. mg/L.
Precision:
|
|
Intra-assay
(n=20 ) | ||
|
Mean
(mg/dl) |
12.4 |
27.3 |
83.5 |
|
SD |
0.28 |
0.40 |
1.61 |
|
CV |
2.25 |
1.48 |
1.93 |
|
|
Inter-assay(n=20) | ||
|
Mean
(mg/dl) |
12.4 |
27.3 |
83.5 |
|
SD |
0.28 |
0.56 |
2.13 |
|
CV |
2.28 |
2.06 |
2.55 |
.
From labcare diagnostic.
Kinetic.
Packagesize:1x12 ml.
No. of tests: 12 tests.
Linearity: From detection limit of 5 U/L to linearity limit of 250 U/L. If the results obtained were greater than linearity limit, dilute the sample 1/10 with NaCl 9 g/L and multiply the result by 10.
From labcare diagnostic.
Fixed time.
Package size: 1x50 tests.
No. of tests: 50 tests.
Linearity: Up to 150 mg/L under the described assay conditions.
From labcare diagnostic.
Kinetic.