From Pshtaz Teb Zaman diagnostic.
Package size: 96 tests.
No. of tests: 96 tests.
Linearity:
|
|
|
Recovery
(%) | |||
|
no |
25-OH-Vitamin-D
undiluted specimen (mlU/L) |
1:2 |
1:4 |
1:8 |
1:16 |
|
1 |
18.3 |
110 |
108 |
* |
* |
|
2 |
37.3 |
92 |
93 |
97 |
* |
|
3 |
58.9 |
98 |
111 |
92 |
89 |
|
4 |
95 |
103 |
106 |
108 |
115 |
|
5 |
112 |
104 |
103 |
99 |
116 |
Sensitivity: The minimum detectable concentration of 25-OH-Vitamin-D was determined by the optical density (OD) of zero standard minus three times the standard deviation (SD). The sensitivity of 25-OH-Vitamin-D kit is 3.5 ng/ml and values less than this amount should be reported as <3.5 ng/ml.
Precision:
intra-assay:
|
Specimen
no. |
No
of tests performed |
Means
(ng/ml) |
SD
|
CV
% |
|
1 |
10 |
12 |
1.7 |
14.2 |
|
2 |
10 |
54 |
2.21 |
4.1 |
|
3 |
10 |
104 |
2.28 |
2.2 |
Inter-assay:
|
Specimen
no. |
No
of tests performed |
Means
(ng/ml) |
SD |
CV
% |
|
1 |
24 |
10.26 |
1.71 |
16.7 |
|
2 |
24 |
33.26 |
3.66 |
11 |
|
3 |
24 |
112.00 |
3.80 |
3.4 |