(App::isLocale('en')) TSH

From Pshtaz Teb Zaman diagnostic.


Package size: 96 tests.


No. of tests: 96 tests.


Linearity:

 

 

Recovery (%)

no

hTSH undiluted specimen (mlU/L)

1:2

1:4

1:8

1:16

1

2.7

100

98

102

96

2

8.4

99

103

99

96

3

16.7

99

94

104

105

4

27.8

97

105

94

85

 


Sensitivity: The minimum detectable concentration of TSH was determined by adding three standard deviations to the mean optical density values of replicated zero standard. The minimum detectable dose using a standard curve is 0.1 mIU/L.


Precision:

          intra-assay:

Specimen no.

No of tests performed

Means (mlU/L)

SD (μg/ml)

CV %

1

16

0.25

0.015

6

2

16

4.35

0.17

3.9

3

16

11.7

0.53

4.5

Inter-assay

For more information.


Specimen no.

No of tests performed

Means (mlU/L)

SD (μg/ml)

CV %

1

20

0.25

0.02

8.0

2

20

4.4

0.31

7.0

3

20

12.1

1.02

8.4