(App::isLocale('en')) T4

From Pshtaz Teb Zaman diagnostic.


Package size: 96 tests.


No. of tests: 96 tests.


Linearity:

 

 

Recovery (%)

no

hT4 undiluted specimen (μg/ml)

1:2

1:4

1:8

1:16

1

8.5

95

106

91

-

2

12

94

91

109

-

3

21

96

90

92

91

4

28

90

94

109

107

 


Sensitivity: The minimum detectable concentration of T4 was determined by adding three standard deviations to the mean optical density values of replicated zero standard. The minimum detectable dose using a standard curve is 0.1 μg/dcurve is 0.1 ng/ml.


Precision:

          intra-assay:

Specimen no.

No of tests performed

Means (μg/ml)

SD (μg/ml)

CV %

1

24

3.6

0.18

5.0

2

24

9.5

0.55

5.7

3

24

16

0.58

3.6

 

Inter-assay

For more information.

Specimen no.

No of tests performed

Means (μg/ml)

SD (μg/ml)

CV %

1

10

3.

0.21

5.5

2

10

9.9

0.76

7.6