From chema diagnostic.
Kinetic.
Available package size: 6 x 10 ml.
No. of tests: 60 tests.
Linearity: The method is linear up to 2000 U/I.
Precision:
|
intra-assay (n=10) |
mean (U/I) |
SD (U/I) |
CV% |
|
sample 1 |
46.21 |
1.80 |
2.18 |
|
sample 2 |
101.64 |
1.01 |
1.77 |
|
intra-assay (n=20) |
mean (U/I) |
SD (U/I) |
CV% |
|
sample 1 |
46.35 |
1.31 |
2.82 |
|
sample 2 |
101.64 |
1.03 |
1.01 |