(App::isLocale('en')) Phosphorus

From chema diagnostic.


Endpoint.


Available package size: 2x 50 ml.


No. of tests: 100 tests.


Linearity: the method is linear up to 20 mg/dl.


Precision:

intra-assay (n=10)

mean (U/l)

SD (U/l)

CV%

sample 1

3.50

0.05

1.50

sample 2

5.87

0.11

1.90

 

intra-assay (n=20)

mean (U/l)

SD (U/l)

CV%

sample 1

3.41

0.08

2.40

sample 2

5.84

0.11

1.90

For more information