From chema diagnostic.
Fixed time.
Available package size: 4 x 125 ml.
No. of tests: 500 tests.
Linearity: the method is linear up to 20 mg/dl.
Precision:
|
intra-assay (n=10) |
mean (mg/dl) |
SD (mg/dl) |
CV% |
|
sample 1 |
1.25 |
0.03 |
2.60 |
|
sample 2 |
3.87 |
0.07 |
1.90 |
|
intra-assay (n=20) |
mean (mg/dl) |
SD (mg/dl) |
CV% |
|
sample 1 |
1.31 |
0.04 |
2.90 |
|
sample 2 |
3.80 |
0.14 |
3.80 |